Abstract : Lag and residual contrast have been quantified in an amorphous selenium activematrix flat-panel imager as a function of frame time, kilovoltage (kV) and megavoltage(MV) x-ray photon energy and amount of incident radiation. The detector contains a 200pm thick a-Se layer deposited on a thin film transistor array of size 8.7 x 8.7 cm2 with an85 pm pitch..